Navigation:  ECN Analysis 1.0 Manual > Selection of processing parameters >

FFT Length

Previous pageReturn to chapter overviewNext page

The 'FFT Length' parameter determines the length of each segment used to obtain all the time-dependent quantities. Longer FFT length will generate spectra extending towards lower frequencies, but the time resolution of the processed data will be reduced because more points are used to generate a single result. Ideally, to generate reliable values of charge and frequency of corrosion events, the 'FFT length' should be sufficiently long that a plateau is visible in the low-frequency end of current and potential power spectral densities spectra.




Fig 13.0 The selection of FFT length can be made in the preference menu.The default length is 1024